Mesin Anesthesia Acoma VIGOR


Specification      
• Power supply (only for DX type)      
Rated power voltage : AC100-240V (as specified)  
Power frequency : 50/60Hz  
Power input : 7VA  
Type of protection against electric shock : Type : Class I
    Level : B type equipment
Auxiliary power outlet : Four; Maximum rated amperage 5A (w/Circuit breaker)
• Anesthetic circuit      
Gas supply pressure : 350~500kPa (common to O2, N20)
Flowmeter : Flow ranee continuously variable; 0- 10 L/min. 
    (common to O2,N2O)  
    (DXonly... w/pressure gauge for each gas, 0-10 x 100kPa)
Canister (CO2 absorption device) capacity : 1,340mL (1,200g)  
Oxygen flush  : 35~75L/min.  
Auxiliary cylinder yoke  : O2 only (one); (DX onh) w/pressure gauge (0-200 x 100kPa)
  Monitors      
Circuit pressure gauge  : -10 - 70hPa  
Oxygen monitor (DX only)  : Galvanic cell measuring system; Button switch
    ON/OFF, and ON by microswitch (as soon as O2 gas flow 
    rises to over 1.8L/min.'. monitor sets to ON)
Measuring ranse/aceuracy : 0~99%O2; 15-25%... ±1 %02; 25-100%...
    ±3%02  
Calibration : 21% calibration  
O2 sensor life : 900.000%H  
#  Safety devices/Alarms      
Ensured minimum O2 flow : Controlled so that when O2 gas flow is less than '1.8 L/min.'. 
    N2O gas won't flow.  
O2 gas supply pressure : If pressure drops below '240~260kPa', whistle sounds.
    Below ' 170-21 OkPa', whistle' sounds at the same tune as 
    N2O supply is shut off.  
Insufficient O2 cone' alarm (DX only)  : O2 cone, selectable at '25%' or '18%',
    and if cone, goes below selected value, alarm activates. 
    (The 'O2 cone. indicator display changes from green to 
    blinking in red. and alarm sounds.)
Calibration unavailable' alarm (DX only) : When O2 sensor is dead or defective, green'--' mark is 
    displayed in the O2 cone, indicator, and calibration 
    unavailable' 'indicator blinks as alarm sounds.
O2 sensor cable disconnected' alarm  : Activates if the cable is disconnected or damaged. 
(DX only)   (The red'—' mark blinks on the O2 cone. indicator display, 
    and alarm sounds.)  
Alarm pause (DX only) : Usable with die following alarms: 
    (1) O2 cone, below '25%'   
         ('Pause' not available when alarm activates at '18%' level.);
    (2) Calibration unavailable; and, 
    (3) O2 sensor cable disconnected.
       
       
    * During the pause, the indicator blinks, and once the condition
       has beencorrected, alarm is automatically cancelled.
# Dimensions/weight      
Outer dimensions...  : 486(w) x 582(d) x 1,360(h)mm;   
Monitor shelf...  : 410(w)x 300 (d)mm, approx. 20kg monitors, etc. mountable; 
Table top...  : 480(w) x220(d)mm at 880mm height;
Main unit weight...  : DX - approx. 70kgs.,   
    ST- approx. 65kgs.  
       
#  Usage conditions      
Room temperature...  : 10~40°C:   
Relative humidity...  : 30-85%;   
Atmospheric pressure...  : 700-1,060hPa  
       